What It Means
Probe Drift / Unstable Readings
Likely Causes
- Probe aged out
- Probe stored dry
- Grounding/bonding noise
- Dirty probe tip
- Air bubbles in flow cell
Next Checks
- Clean probe and recalibrate with fresh standards
- Remove bubbles and confirm steady sample flow
- Check bonding/grounding and nearby electrical noise
- Replace probe if it cannot hold calibration
Verify Before Repair
SplashLens is a field reference, not a repair guarantee. Confirm the visible code, exact model, and current manufacturer procedure before replacing parts, bypassing safeties, or quoting a customer.
SplashLens is independent and is not affiliated with Commercial Chem Controllers or other equipment manufacturers.